Portable Meter (SEM-103)
Product Code: SEM-103
Availability: In Stock

Comment(s) / Special Request(s):

Product Portable Meter (SEM-103)
Model SEM-103
Measuring Gas Oxygen Carbon Monoxide Nitrogen monoxide Flue gas temperature
Detection Method Galvanic Cell Chronoamperometry Chronoamperometry Thermocouple (type k)
Measuring Range 0.0~22.0vol% 0~2,000ppm 0~2,000ppm 0~1,200˚C
Resolution 0.1vol% 1ppm 1ppm -
Indication Accuracy ±0.7vol%
(0.0~22vol%)
0~300ppm ±15%
301~2,000ppm ±10%
0~300ppm ±15%
301~2,000ppm ±10%
-
Alarm Method LCD
Response Time Within 15 seconds at 90% response Within 20 seconds at 90% response Within 20 seconds at 90% response -
Power Supply 4 x D size alkaline dry battery or AC100V (AC adaptor)
Battery Life Alkaline dry batteries: 15 hours
Operating Temperature -10~40˚C; below 90%RH (non-condensing)
Dimensions 300(W) X 203(H) X 195(D) mm (without drain tap)
Weight 4 kg
Standard Accessory Nox removal filter, water stopper, sampler, dust filter, connecting hose, tedlar bag and uni tube for calibration, temperature sensor adaptor, temperature sensor, carrying case, AC adpator
Additional Info -

 

Product Portable Meter (SEM-103)
Model SEM-103
Measuring Gas Gas temperature at sensor Carbon dioxide Combustion efficiency
Detection Method Thermistor Caliculation Caliculation
Measuring Range -20~60˚C 0.0~21.0% 0.0~100.0%
Resolution - 0.10% -
Indication Accuracy ±5˚C ±2˚C  
Alarm Method LCD    
Response Time Within 15 seconds at 90% response Within 20 seconds at 90% response Within 20 seconds at 90% response
Power Supply 4 x D size alkaline dry battery or AC100V (AC adaptor)
Battery Life Alkaline dry batteries: 15 hours
Operating Temperature -10~40˚C; below 90%RH (non-condensing)
Dimensions 300(W) X 203(H) X 195(D) mm (without drain tap)
Weight 4 kg
Standard Accessory Nox removal filter, water stopper, sampler, dust filter, connecting hose, tedlar bag and uni tube for calibration, temperature sensor adaptor, temperature sensor, carrying case, AC adpator